Main Specification of Anesthesia Workstation :
Main body : High-strength engineering plastic rack, light, beautiful and corrosion resistant
Scope of application : Adult and Child
Gas source : O2: 0.28~0.6MPa N2O: 0.28~0.6MPa
Flow meter : O2: 0.05~1.0L/min 1.1~10L/min
N2O: 0.05~1.0L/min 1.1~10L/min
O2, N2O linkage and N2O Stopper
When using nitrous oxide, oxygen concentration>25%;
When the oxygen pressure<0.2Kpa, the flow of nitrous oxide would be cut off.
Flow rate of rapid oxygen supply : 25~75L/min
Low oxygen pressure alarm : There will be sound alarm when the oxygen pressure < 0.2MPa
Vaporizer : It has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0~5 vol%.
Among Halothane,Enflurane,Isoflurane and Sevoflurane, two can be chosen for application as required by the customer.
Respiratory circuit working mode : All-close, semi-close, semi-open APL ≤12.5kpa
Respiratory Bellows : Bellows for adults, bellows for children tidal volume range:0~1,500 ml
Main Specification of Ventilator
Display mode : High-definition 5.7” LCD screen display
Ventilation mode : IPPV, SIPPV, SIMV, MANUAL
Ventilation function : PEEP,SIGH, IRV,
Ventilation parameters : Tidal volume 50~1500 ml Rate 2~99 bpm
SIMV rate 2~20 bpm I:E 2:1~1:8
Inspiratory trigger pressure -10~20cmH2O PEEP 2~20cm H2O Pressure Range5~60cmH2O
SIGH1.5 times the inspiratory time (60-120 adjustable)
Parameters for ventilation monitoring tidal volume, minute ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, average pressure, pressure - time waveform, flow rate - time
waveform, PEEP, inspiratory trigger pressure
Security Alarm System
Oxygen Concentration Alarm : Upper limit setting range (21%~99%)
Low limit setting range (18%~80%)
Airway Pressure Alarm : Upper limit setting range 5~60 cm H2O
Low limit setting range 0~50 cmH2O
Minute Ventilation Volume Alarm : Upper limit setting range 1~20 L/min
Low limit setting range 0~19L/min
Tidal Volume : Upper Limit Setting Range 100~1500ml
Low Limit Setting Range : 0~1400ml
Sustained high-pressure alarm : It will give alarm when stress have consistently been higher than 2.5cm H2O
(Suffocation alarm, Power alarm, Intubation off alarm, battery voltage alarm)
Power : AC 220V 50Hz
(Mechanical Arms End-tidal CO2 monitoring)
We are Manufacturer, Exporters and Suppliers of Anaesthesia Workstation Gold at wholesale prices from India.
Model No :
APAW 111 Anaesthesia Workstation Gold